GENOSS DES

Sirolimus Eluting Coronary Stent System


· 70µm of Co-Cr strut thickness
· Bioresorbable polymer abluminal coating (thickness < 4µm)
· Sirolimus dose 1.15µm / mm²



70µm of Co-Cr strut thickness


Bioresorbable polymer abluminal coating
(thickness < 4µm)


Sirolimus
1.15 µm / mm²

• Product Details

Crossing profile

Polymer coating integrity after expansion

Angiographic and IVUS findings (9-month follow up)

GENOSS DES (n=38) B Company (n=39) p value
In-Stent late lumen loss (mm) 0.11±0.25 0.16±0.43 0.67
In-Segment late lumen loss (mm) 0.11±0.26 0.15±0.43 0.56
IVUS lumen CSA (mm2 ) 0.69±1.44 0.59±0.81 0.70
IVUS stent CSA (mm2 ) 0.10±0.70 0.26±0.42 0.25
EEM (external elastic membrane) -0.24±0.79 0.49±1.32 0.006

Clinical cases

Clinical case 1
Clinical case 2
Clinical case 2
Clinical case 2